Laminar Flow Meter |
|

The Laminar Flow Meter is a highly precise, differential pressure flow meter that utilizes the phenomenon called the Hagen-poiseuille law, expressed in terms of the ratio showing the loss of pressure in a volumetric flow rate of fluids flowing through a cylinder or a dihedral device over a fixed distance.
|
Special characteristics |
- Has a wide range of measurement and is linear for dealing with the relationships between differential pressure and flow volume.
- Calibrated with standard sonic nozzles for high precision.
- Minimal pressure loss.(max 2.6kPa)
- Superb linearity and reproducibility.
- Excellent response(max@30Hz)
- Appropriate for measuring microscopic flow volumes.
- Easy to manage.
|
Usage |
- Measurements of the intake air flow in internal combustion engines.
- Flow volume measurements for vacuum pumps, fans, blowers, etc.
- Specific measurements for carburetors, throttle valves, mufflers, etc.
- Specific measurements for various air machinery.
- Measurements of respiratory volume.
- Flow volume measurements for propane and town gas.
- Flow volume regulation during the process.
|
Specifications |
@ @ |
| Flow volume range |
Fetch differential pressure*50`700Pa
general precision:}0.5% |
|
|
l/sec
|
l/min
|
| LFM-1 |
0.013@`@0.2
|
0.8@`@12
|
| LFM-2 |
0.026@`@0.4
|
1.6@`@24
|
| LFM-3 |
0.052@`@0.8
|
3.2@`@48
|
| LFM-4 |
0.104@`@1.6
|
6.3@`@96
|
| LFM-5 |
0.21 @`@3.2
|
12.5@`@192
|
| LFM-6 |
0.42 @`@6.4
|
25.0@`@384
|
| LFM-7 |
0.83 @`@12.8
|
50.0@`@768
|
| LFM-8 |
1.66 @`@25.6
|
100@`@1540
|
| LFM-9 |
3.34 @`@51.2
|
200 `@3070
|
| LFM-10 |
6.7 @`@102
|
402@`@6140
|
| LFM-11 |
13.4 `@205
|
804@`@12300
|
| LFM-12 |
27.0@`@410
|
1620@`@24600
|
|
|
 |